WQM-701

Compact Multi-Parameter Water Quality Analyzer

Wall-mounted analyzer measuring free chlorine or ClO2, turbidity, pH, conductivity and temperature in one compact, reagent-free unit.

Multi-parameterFree chlorineTurbiditypHConductivityReagent-freeRS485Wireless upload
WQM-701 Compact Multi-Parameter Water Quality Analyzer
Water Quality Monitoring
Family
WQM Series
Category
Water Quality Monitoring
Parameters
Free chlorine / ClO2, turbidity, pH, temperature, conductivity (optional)
Display
7-inch colour touchscreen

01Overview

Engineered for the field

A compact water quality analyzer built for secondary water supply and distribution networks, measuring the key drinking water safety parameters in one wall-mounted enclosure: free chlorine or chlorine dioxide, turbidity, pH, temperature and optional conductivity. Measurement is reagent-free, the pH electrode is the only consumable, and self-cleaning turbidity and chlorine sensors extend service intervals. The turbidity channel resolves below 0.1 NTU for drinking water, while a membrane-free platinum chlorine electrode with flow stabilisation and pH and temperature compensation copes with zero chlorine, intermittent flow and low pressure. At just 30 × 40 × 18 cm, with a 7-inch touchscreen, RS485 and wireless data upload, it suits unattended pump rooms and rural water stations.

02Key Features

What makes the WQM-701 different

01

Five parameters, one unit

Free chlorine or ClO2, turbidity, pH, temperature and optional conductivity in a single enclosure.

02

Low running cost

Reagent-free measurement with the pH electrode as the only consumable.

03

Drinking water sensitivity

Turbidity below 0.1 NTU and compensated chlorine readings to guide disinfectant dosing.

04

Robust chlorine electrode

Membrane-free platinum electrode tolerates zero chlorine, intermittent flow and low pressure.

05

Remote supervision

RS485 and wireless data upload bring dispersed sites into one monitoring view.

03Technical Specifications

Detailed specifications

Overview

Parameters
Free chlorine / ClO2, turbidity, pH, temperature, conductivity (optional)
Display
7-inch colour touchscreen
Communication
RS485 Modbus RTU, wireless data upload
Dimensions
30 × 40 × 18 cm (W × H × D)

Measurement

Turbidity
90° scattering; 0 to 20 NTU (0 to 100 NTU optional); 0.001 NTU resolution; ±2% or ±0.02 NTU
Free chlorine / ClO2
Three-electrode amperometric; 0 to 2 mg/L (0 to 5 mg/L optional); 0.001 mg/L resolution; ±5% or ±0.03 mg/L
pH
Electrochemical; 0 to 14 pH; 0.01 pH resolution; ±0.1 pH
Conductivity (optional)
Four-electrode; 0 to 10 mS/cm (0 to 100 mS/cm optional); ±1.5% FS
Temperature
NTC10K; -5 to +60 °C; ±0.2 °C
Response time
≤0.5 min (turbidity, pH, conductivity); ≤2 min (chlorine)

Sample

Sample temperature
5 to 40 °C
Sample pressure
1 to 4 bar
Sample flow rate
Turbidity 320 to 1000 mL/min; other parameters over 180 mL/min

Power and Environment

Power supply
85 to 264 V AC, 50/60 Hz
Power consumption
15 W average, 35 W max.
Operating temperature
Controller -10 to +50 °C; sensors -15 to +55 °C
Relative humidity
Under 90%, non-condensing
Weight
10 kg
Installation
Wall or panel mounted, flow-through sampling

04Applications

Built for these use cases

  • Use case 01

    Secondary water supply pump stations

  • Use case 02

    Municipal distribution networks

  • Use case 03

    Township and village water plants

  • Use case 04

    Industrial raw water monitoring

Industries

Deployed across these sectors

05Resources

Datasheets, manuals and technical documents

Documents are shared on request. Tell us which file you need and our team will send it to you.

  • WQM-701 Datasheet

    Datasheet

    Request
  • WQM-701 User Manual

    User Manual

    Request

06Related products

Complementary engineering